Flownetix uses cookies on your computer to help us make this website better. To find out more about the cookies, see our privacy policy. 

FAQ  

Q: How do you ensure continued performance of the 300series products over their lifetime?

A: The 300series incorporates precision oscillators and analog references. All electronic functions are tested and compared to these references every 2 seconds and used to correct any measurements. Any changes due to time, temperature and component drift are automatically corrected for on a continuous basis. The amount of correction is available from the meter diagnostics at any time and can be checked to ensure operation within allowable limits.

Click here to return to the list of questions.